Title | Temperature and temporal dependence of neutral density transmittance standards |
Publication Type | Journal Article |
Year of Publication | 2012 |
Authors | Koo, A., and Hamlin J.D. |
Journal | Metrologia |
Volume | 49 |
Issue | 2 |
Pagination | S68 - S72 |
Date Published | 2012 |
ISSN | 00261394 (ISSN) |
Keywords | Engineering, Manufacture, Metallurgical engineering, Neutral densities, Spectral transmittance, temperature dependence, Temporal dependence |
Abstract | The Schott series of NG glasses are frequently used to manufacture neutral density transmittance standards for validation of spectrophotometer systems as well as for comparisons of regular spectral transmittance scales. A study has been made of the temperature and temporal dependence of transmittance in these types of filters. The temperature dependence of transmittance is found to scale as -ln(T). The filter transmittance was found to vary significantly with time shortly after manufacture but appears to be stabilizing nine months after beginning measurements. © 2012 BIPM & IOP Publishing Ltd. |
URL | http://www.scopus.com/inward/record.url?eid=2-s2.0-84857989474&partnerID=40&md5=c5320d4d05d3a47907592cdb52b7062d |
DOI | 10.1088/0026-1394/49/2/S68 |