Callaghan Innovation Research Papers

Back to Research Papers

TitleResistance network for verifying the accuracy of resistance bridges
Publication TypeJournal Article
Year of Publication1997
AuthorsWhite, D.R., and Williams J.M.
JournalIEEE Transactions on Instrumentation and Measurement
Volume46
Issue2
Pagination329 - 332
Date Published1997
ISSN00189456 (ISSN)
KeywordsBridge circuits, Calibration, Cryogenic current comparators, Cryogenics, Electric resistance measurement, Resistors
AbstractA novel resistance network for verifying the accuracy of resistance bridges is described. The network of four resistors generates 35 four-terminal resistance values all interrelated by the formulae for series and parallel connections of resistors, and by appropriate choice of the resistors different resistance scales can be generated. Although many resistance bridges have internal calibration procedures, an independent check of the accuracy of such bridges when measuring actual resistances is often desirable. The network provides such a check and can be used to verify the accuracy of resistance bridges on a regular basis. It has been used to assess the accuracy of a cryogenic current comparator bridge and demonstrates an accuracy approaching 1 nΩ/Ω at 100 Ω. © 1997 IEEE.
URLhttp://www.scopus.com/inward/record.url?eid=2-s2.0-0031121525&partnerID=40&md5=1ea1059cace1c12168b89b3c2a94395a
DOI10.1109/19.571849

Back to top