TitleImpedance bridge ratio transformer calibration by reference-step conductance build-up
Publication TypeConference Paper
Year of Publication2004
AuthorsConvey, A.C.
Conference NameCPEM Digest (Conference on Precision Electromagnetic Measurements)
Date Published2004
KeywordsImpedance bridges, Input voltages, Uncertainty analysis, Voltage dividers
AbstractA reference-step conductance build-up method suitable for automation has been developed for calibrating the ratio dial transformer of a wide-range impedance bridge. The uncertainties over the range 40 Hz to 20 kHz are estimated not to exceed 0.2 μV/V referred to the divider input voltage. At 100 kHz, the uncertainties do not exceed 2 μV/V. ©2004 IEEE.