Title | DC performance of a high accuracy thermal transfer standard |
Publication Type | Conference Paper |
Year of Publication | 2010 |
Authors | Early, M.D., Garcocz M., and Christian L.A. |
Conference Name | CPEM Digest (Conference on Precision Electromagnetic Measurements) |
Date Published | 2010 |
Keywords | Ac voltage, Common mode, DC performance, Direct impact, Electromagnetism, High-accuracy, Thermal transfer |
Abstract | The dc performance of a widely-used high-accuracy commercial thermal transfer standard is found to be significantly affected by a variable contact resistance on its lower impedance ranges. This has a direct impact on the ac voltages measured by the instrument on these ranges. Cleaning the contacts and ensuring clearance around an input common mode choke effectively resolves the problem. ©2010 IEEE. |
URL | http://www.scopus.com/inward/record.url?eid=2-s2.0-77956620856&partnerID=40&md5=4b8d08ae7c4e95c9d43fd590d99e4046 |